Handy, fast and precise - three of the reasons that make the Alcotest 7410 (Plus RS) to become the world standard in hand-held breath-alcohol measuring instruments. The long-term stable DrägerSensor gives fast exact results on an illuminated display with three digits.
Characteristics
Measuring principle: Electrochemical sensor
Measuring units: mg/L, ‰, % (further measuring units available)
Measurement range: 0.00 to 1.50 mg/L; 0.00 to 3.00 ‰; 0.000 to 0.300 %
Display range: 0.00 to 9.99 mg/L; 0.00 to 9.99 ‰; 0.000 to 0.999 %
Ambient conditions For operation: – 5 to 45°C /23 to 113°F (and at – 20°C/– 4°F but not for more than 30 min and only after being kept above 0°C/32°F until use) For storage: – 40 to 65°C/– 40 to 149°F
Sampling: Blowing time: 4 to 12 s (depends on the intensity) Breath flow: > 6 L/min Breath volume: >1.2 L (adjustable)
Measurement accuracy (max. error of repeatability with an ethanol standard): 0 to 0.5 mg/L: ± 0.03 mg/L 0 to 1.0 ‰: ± 0.05‰ > 0.5 mg/L: ± 5 % of measured value >1.0 ‰:
Sensitivity drift: typically 0.6 % of measured value/month
Intervals before readiness for operation: 1st measurement approx. 15 s after determining a previous test result in range: up to 0.19 mg/L: 10 s > 0.95 mg/L: 90 s
Calibration interval recommended 6 months (depends on required accuracy)